APPLICATION :
Used for produce clean air referring to cGMP Standard (Class A/Class 100, Class B/Class 100 &Class C/Class 100.000)
A. LAMINAR AIR FLOW BENCH FOR MICROBIOLOGY
Used for examination of microbiology, vaccine, etc
SPECIFICATION :
Working Zone |
: |
± 90 or 120 x 60 x 50 cm (WxDxH) (or as customer required) |
Class |
: |
100 FDE STD 209 E / Class A CPOB / CGMP 2006 |
Material |
: |
Stainless Steel AISI 304 |
Filtering Element |
: |
Pre filter (non woven, washable, eff. ± 40%), Hepa Filter (Eff. 99,999% H14) |
Blower |
: |
EBM Zeihl, Germany |
Lighting |
: |
Tube Lamp and UV Lamp |
Pressure Gauge |
: |
Dwyer Minihelic Differential Pressure Gauge |
Instrumentation |
: |
switch for blower, TL, TUV and hour counter |
Door Chamber |
: |
Polycarbonate door with side glasses, vertical sliding door |
Documentation |
: |
DQ, IQ, OQ dan PQ Documentation |
|
|
Individual Certificate for Hepa filter |
|
|
Particle Test and Air Velocity |
LAMINAR AIR FLOW BENCH FOR MIKROBIOLOGY
B. LAMINAR AIR FLOW WEIGHING/SAMPLING HOOD
SPECIFICATION :
Working Zone |
: |
± 200 x 180 cm (or as customer required) |
Class |
: |
100 FDE STD 209 E / Class A CPOB / CGMP 2006 |
Material |
: |
Stainless Steel AISI 304 |
Filtering Element |
: |
Pre filter (non woven, washable, eff. ± 40%), Hepa Filter (Eff. 99,999% H14), Medium Filter (Eff. ± 95%) |
Blower |
: |
EBM Zeihl, Germany |
Lighting |
: |
Tear Drop Lamp |
Pressure Gauge |
: |
Dwyer Minihelic Differential Pressure Gauge |
Curtain |
: |
PVC Curtain blueclear, antistatic |
Installation |
: |
Knock Down System |
Optional |
: |
AISI 304 Table for Weghing |
|
|
Pass Through with Roller Conveyor |
Documentation |
: |
DQ, IQ, OQ dan PQ Documentation |
|
|
Individual Certificate for Hepa filter |
|
|
Particle Test and Air Velocity |
LAMINAR AIR FLOW WEIGHING & SAMPLING HOOD
C. LAMINAR AIR FLOW FOR PRADUCTIO OF PARENTHERAL SOLUTION, INFUS, OPTHAMIC SOLUTION, POWDER INJECTION, AND OTHER PURPOSE